Model Number PCB00
Device Problems Break (1069); Deformation Due to Compressive Stress (2889)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Age/date of birth: unknown/not provided.Sex/gender: unknown/not provided.Date of event: unknown/not provided.If implanted, give date: not applicable, as the lens was not implanted.If explanted, give date: not applicable, as the lens was not implanted; therefore, it was not explanted.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that an intraocular lens (iol) was defective.The tip of the cartridge looked melted and deformed.Patient involvement is unknown, however, there was no surgical intervention required.The patient is doing fine post-operatively.No additional information was provided.
 
Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Manufacturer Narrative
Device evaluation: the device was not returned at the manufacturing site; therefore; product testing could not be performed, and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.A search of complaints revealed two other complaints were received under this production order; however, product deficiency was not identified.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Type of Device
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key9257829
MDR Text Key166360674
Report Number2648035-2019-01157
Device Sequence Number1
Product Code HQL
UDI-Device Identifier05050474558328
UDI-Public(01)05050474558328(17)220625
Combination Product (y/n)Y
PMA/PMN Number
P980040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup,Followup
Report Date 01/01/2005,11/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/25/2022
Device Model NumberPCB00
Device Catalogue NumberPCB0000230
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received10/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-