Model Number URF-V
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/09/2019
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation. The exact cause of the reported event could not be conclusively determined at this time. If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp. (omsc) was informed that as a result of routine microbiological testing by the user facility, the sample collected from the instrument channel of the subject device tested positive for the following microbes. Klebsiella pneumoniae. Escherichia coli. Enterococcus faecalis. Staphylococcus aureus. Providencia rettgeri. Myroides sp. The total number of these microbes is >100 cfu/ml. The device had been reprocessed with an olympus automated endoscope reprocessor model etd4 (not available in the usa) using peracetic acid. There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information. The subject device has not been returned to omsc but was returned to olympus europa (b)(4). (b)(4) investigated the subject device and found that there was no malfunction. (b)(4) (b)(4) sent the device to a third party laboratory for microbiological testing. As a result of the testing, no microbe was detected from the sample collected from the distal end and the instrument channel of the device. The exact cause of the reported event could not be conclusively determined. If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameURETERO-RENO VIDEOSCOPE
Type of DeviceURETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key9258338
MDR Text Key219774239
Report Number8010047-2019-03762
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K072957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation
Type of Report Initial,Followup
Report Date 11/25/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberURF-V
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received11/07/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-