• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problem Microbial Contamination of Device (2303)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/01/2019
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of microbiological testing by the user facility, the sample collected from the subject device tested positive for staphylococcus saprophyticus.The device had been manually reprocessed using peracetic acid.There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to omsc but was returned to olympus (b)(4).(b)(4) sent the device to a third party laboratory for microbiological testing.As a result of the testing, no microbe was detected from the sample collected from the all channels of the device.The testing result cleared the (b)(4) guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key9259235
MDR Text Key172831943
Report Number8010047-2019-03773
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/09/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-