COOK ENDOSCOPY MEMORY II DOUBLE LUMEN EXTRACTION BASKET; FFL, DISLODGER, STONE, BASKET, URETHRAL, METAL
|
Back to Search Results |
|
Model Number G25149 |
Device Problem
Fracture (1260)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/18/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation evaluation: our laboratory evaluation of the product said to be involved determined the drive wire separated from the handle.The device was returned with the basket fully retracted into the sheath.During a functional test the handle was manipulated and there was no response from the basket.The handle was disassembled and it was noted that the drive wire had separated from the handle.Under magnification it was noted there was nesting inside the purple hub.For further evaluation of the drive wire cable and basket, the catheter was cut to push the drive wire cable out of the sheath.The basket was fully formed and intact.Solder was present on the handle cannula at the joint.No other anomalies were detected with the device.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Basket advancement/retraction difficulties and nesting of the drive wire can occur if the device experiences excessive pressure.Resistance in basket movement and bends in the catheter can occur if the elevator of the endoscope is used to deflect the device at a sharp angle.The instructions for use indicate: "advance device through channel, in short increments, until basket sheath exits endoscope." the instructions for use state: "confirm desired position of basket sheath relative to target.Advance basket out of sheath.Caution: pulling on sheath while advancing or retracting basket may damage device, rendering it inoperable." prior to distribution, all memory ii double lumen extraction baskets are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
Event Description
|
During an endoscopic procedure for removal of bile stones, the physician used a cook memory ii double lumen extraction basket.The basket was advanced over a wire guide and the user attempted to open the basket to capture the 5 mm stone but the basket would not open.Another of the same device was used instead.There was no reportable information at this time.The device was evaluated on 09-oct-2019 and the drive wire had separated from the handle.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Search Alerts/Recalls
|
|
|