MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HD SYS. CDX W/BIBAG BLUE STAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HD SYS. CDX W/BIBAG BLUE STAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 191126
Device Problems Thermal Decomposition of Device (1071); Melted (1385)
Patient Problem No Patient Involvement (2645)
Event Date 10/15/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had low temperature.A fresenius field service technician (fst) replaced the heater fuse.A picture of the heater fuse was provided and there were visual indications of a burned and melted cable.The biomed stated that they did not observe any burning smell, smoke, spark, flame, arcing, or any other visible heat or electrical damage related to the burned and melted cable found on the heater fuse.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The machine has approximately 14-15 hours and the heater fuse was the original fresenius part on the machine.The biomed confirmed that the machine has not had any past problems with failing the electrical leakage test.The biomed stated that the machine is plugged into a hospital grade ground-fault circuit interrupter (gfci) outlet.The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the burned and melted heater fuse.The heater fuse was replaced, which resolved the issue.The unit was returned to service at the user facility without issue and without reoccurrence of the event.The distribution board was discarded by the biomed and is not available to be returned to the manufacturer for physical evaluation.
 
Manufacturer Narrative
Plant investigation: the heater rod was returned to the manufacturer for physical evaluation.An exterior visual inspection of the returned heater rod showed the wire sheath was melted, and the connector was discolored.The heater rod was connected to a test machine and ran through rinse mode, dialysis mode, and self-test.The heater rod passed self-test.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.An on-site evaluation was performed by a fresenius field service technician (fst).The fst resolved the low temperature issue by replacing the heat-damaged heater fuse cable component.The machine was tested and passed functional checks.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was determined to be the heat damaged heater fuse.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HD SYS. CDX W/BIBAG BLUE STAR
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key9267560
MDR Text Key164656478
Report Number2937457-2019-03300
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861102099
UDI-Public00840861102099
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173972
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 01/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2019
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number191126
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2019
Is the Reporter a Health Professional? Yes
Device Age MO
Date Manufacturer Received01/06/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-