CONCORD MANUFACTURING 2008T HD SYS. CDX W/BIBAG BLUE STAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Catalog Number 191126 |
Device Problems
Thermal Decomposition of Device (1071); Melted (1385)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/15/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
|
Event Description
|
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had low temperature.A fresenius field service technician (fst) replaced the heater fuse.A picture of the heater fuse was provided and there were visual indications of a burned and melted cable.The biomed stated that they did not observe any burning smell, smoke, spark, flame, arcing, or any other visible heat or electrical damage related to the burned and melted cable found on the heater fuse.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The machine has approximately 14-15 hours and the heater fuse was the original fresenius part on the machine.The biomed confirmed that the machine has not had any past problems with failing the electrical leakage test.The biomed stated that the machine is plugged into a hospital grade ground-fault circuit interrupter (gfci) outlet.The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the burned and melted heater fuse.The heater fuse was replaced, which resolved the issue.The unit was returned to service at the user facility without issue and without reoccurrence of the event.The distribution board was discarded by the biomed and is not available to be returned to the manufacturer for physical evaluation.
|
|
Manufacturer Narrative
|
Plant investigation: the heater rod was returned to the manufacturer for physical evaluation.An exterior visual inspection of the returned heater rod showed the wire sheath was melted, and the connector was discolored.The heater rod was connected to a test machine and ran through rinse mode, dialysis mode, and self-test.The heater rod passed self-test.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed.
|
|
Manufacturer Narrative
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.An on-site evaluation was performed by a fresenius field service technician (fst).The fst resolved the low temperature issue by replacing the heat-damaged heater fuse cable component.The machine was tested and passed functional checks.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was determined to be the heat damaged heater fuse.
|
|