Catalog Number 1550350-28
Device Problems Material Rupture (1546); Device Dislodged or Dislocated (2923)
Patient Problem Vessel Or Plaque, Device Embedded In (1204)
Event Date 10/12/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents.The investigation determined the reported difficulties and subsequent treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event Description
It was reported the procedure was performed to treat a heavily calcified lesion in the heavily tortuous right coronary artery.The 3.5x28 mm xience sierra stent delivery system (sds) was advanced, however the balloon ruptured during the first inflation to eight atmospheres.The stent dislodged and was successfully crushed to the vessel wall with a non-abbott stent.There was no adverse patient sequela or a clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key9270004
MDR Text Key164838334
Report Number2024168-2019-13132
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648227448
UDI-Public08717648227448
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/29/2020
Device Catalogue Number1550350-28
Device Lot Number9052542
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/14/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-