Product analysis : analysis confirmed display has deep scratches.Replaced display with salvaged material.Salvaged material was inspected and tested per applicable requirements.Programmer powers up with "operating system not found" error.Reloaded software.Confirmed missing "rubber pad" on case.This part does not affect functionality of the programmer and, due to lack of parts available, will not be replaced.Passed all functional, safety and systems tests.If information is provided in the future, a supplemental report will be issued.
|
It was reported that the programmer's touchscreen is not working, and there are deep scratches in the display.In addition, there is a rubber pad missing from the upper right corner.The programmer has been returned for service.No patient complications have been reported as a result of this event.
|