Model Number 29901A
Device Problems No Display/Image (1183); Mechanical Problem (1384); Component Missing (2306)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/05/2019
Event Type  malfunction  
Manufacturer Narrative
Product analysis : analysis confirmed display has deep scratches.Replaced display with salvaged material.Salvaged material was inspected and tested per applicable requirements.Programmer powers up with "operating system not found" error.Reloaded software.Confirmed missing "rubber pad" on case.This part does not affect functionality of the programmer and, due to lack of parts available, will not be replaced.Passed all functional, safety and systems tests.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer's touchscreen is not working, and there are deep scratches in the display.In addition, there is a rubber pad missing from the upper right corner.The programmer has been returned for service.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK ENCORE PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9271489
MDR Text Key167506339
Report Number3004593495-2019-01155
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number29901A
Device Catalogue Number29901A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2019
Date Manufacturer Received11/11/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-