Catalog Number 362780
Device Problem Incorrect, Inadequate or Imprecise Resultor Readings (1535)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6). A device evaluation is anticipated, but has not yet begun. Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that 30 bd vacutainer® cpt¿ nh: ~130 iu ficoll¿: 2. 0ml experienced hemolysis during use. The following information was provided by the initial reporter: cpt, isolated cells contaminated with erythrocytes and low vitality laboratory receives centrifuged cpt from participants of a clinical study; the cells collected above the gel are contaminated with erythrocytes, often the sample looks hemolytic before further processing; total number of cells is high, 20-30 million cells, but a large proportion of them are probably erythrocytes. The vitality of the cells is below 87% (study sponsor gives 90% as a minimum requirement). After freezing (with dmso, protocol dictated by study sponsor), the vitality is> 90% and the cell count is 7-8 million. Study sites centrifuge 17 min between 1650 rcf and 1800 rcf.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameBD VACUTAINER® CPT¿ NH: ~130 IU FICOLL¿: 2.0ML
Type of DeviceBLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key9273518
MDR Text Key191174352
Report Number1917413-2019-02304
Device Sequence Number1
Product Code JCF
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K891407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2019
Device Catalogue Number362780
Device Lot Number8059582
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 11/04/2019 Patient Sequence Number: 1
-
-