MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Death (1802); Hemorrhage/Bleeding (1888)
Event Date 10/24/2019
Event Type  Death  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that a few hours post-operatively to a successful cryo ablation procedure, the patient experienced massive bleeding from the femoral vein.Blood transfusions were administered without resolve as the patient's body began shutting down.Cardiopulmonary resuscitation then took place without success.The patient is deceased.
 
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files for the date of the event showed system notice 50013 ¿refrigerant level is low.¿ clinical issues were encountered during the case.In conclusion, the system notice was not related to the reported adverse event.There is no indication of relation of adverse event to the performance and malfunction of the product.The physical product was not returned for investigation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported the patient's official cause of death was bleeding.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9275479
MDR Text Key165010481
Report Number3002648230-2019-00747
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age58 YR
-
-