MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551NAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551NAS
Device Problem Obstruction of Flow (2423)
Patient Problems Hyperglycemia (1905); Heart Failure (2206)
Event Date 10/21/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that they were hospitalized due to high blood glucose on (b)(6) 2019 at 2pm with blood glucose of 774 mg/dl and stated that the insulin was not getting delivered.The customer experienced symptoms such as heart failure and kidney failure and sickness.The customer was treated with fluids, insulin drip then changed to injections, heart monitoring, dialysis, taking care of diabetes.The customer was wearing the insulin pump during the hospitalization.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and delivery accuracy test at 0.08750 inches.The stop (idle) current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.The no delivery alarm functions properly during the basic occlusion test and occlusion test.No unexpected no delivery alarm noted during testing.Device uploaded properly using carelink.Device had minor scratched display window, broken battery tube threads, scratched reservoir tube window and a partially broken off reservoir tube lip.The test p-cap and reservoir does lock in place in the reservoir compartment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551NAS
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key9276484
MDR Text Key165198378
Report Number2032227-2019-112763
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169503663
UDI-Public(01)00643169503663
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-551NAS
Device Catalogue NumberMMT-551NAS
Device Lot NumberA4551NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2020
Is the Reporter a Health Professional? No
Date Manufacturer Received01/11/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/15/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age42 YR
Patient Weight138
-
-