• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. ARROW CVC KIT: 3-LUMEN 7 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. ARROW CVC KIT: 3-LUMEN 7 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE Back to Search Results
Catalog Number CA-25703-A
Device Problem Entrapment of Device (1212)
Patient Problems No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
Event Date 10/21/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer reports: the guide wire was left in the patient inside of the catheter.There was an intervention as the patient was sent to another department so the device could be removed by being snared.The complete device was removed under ct without incident.The patient condition is reported as ok.
 
Manufacturer Narrative
Qn#(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
The customer reports: the guide wire was left in the patient inside of the catheter.There was an intervention as the patient was sent to another department so the device could be removed by being snared.The complete device was removed under ct without incident.The patient condition is reported as ok.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC KIT: 3-LUMEN 7 FR X 20 CM
Type of Device
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key9279074
MDR Text Key177799070
Report Number9680794-2019-00430
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
PMA/PMN Number
K900263
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/05/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date03/31/2021
Device Catalogue NumberCA-25703-A
Device Lot Number13F19F0543
Was Device Available for Evaluation? No
Date Manufacturer Received12/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-