BOSTON SCIENTIFIC CORPORATION MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
|
Back to Search Results |
|
Model Number 7585 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/12/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by mfr.: the device was returned for analysis.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was blood in the inflation lumen and guidewire lumen.The balloon was loosely folded.There were numerous hypotube kinks throughout the device.There was a complete hypotube separation 20.5cm distal of the strain relief.The hypotube fracture surfaces were ovaled, which suggests the device was kinked prior to separation.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported shaft kink, as there were numerous kinks in the hypotube with a complete hypotube separation resulting from a kink.
|
|
Event Description
|
Reportable based on device analysis completed on 14-oct-2019.It was reported that shaft kink was encountered.A 2.00mm x 15mm maverick 2 balloon catheter was selected for use.However, during procedure, it was noted that the delivery shaft was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed shaft detached/separated.
|
|
Search Alerts/Recalls
|
|
|