|
Model Number 2420-0007 |
Device Problems
Break (1069); Fluid/Blood Leak (1250); Separation Failure (2547)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/14/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Race and ethnicity: white.The devices have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.
|
|
Event Description
|
It was reported when the rn arrived at the patient's bedside found that the iv that was connected to the patient was broken in two.It was found to be broken about 4 inches below the pump with the other half still connected to the patient.The patient at this time was experiencing confusion, acutely disoriented, anxious, and intervening with other forms of treatment such as picking at dressings covering wounds and pulling out the second iv that was not in use.The rn heard the iv pole being jostled prior to entering the room.The patient grabbed the iv tubing and pulled on it hard enough to break it and the medication (heparin) leaked on the floor.Although it was reported that patient care was delayed it did not contribute to, or result in serious adverse impact to patient.
|
|
Manufacturer Narrative
|
The customer¿s report that the tubing broke at junction ¿separated¿ was confirmed during visual inspection.The set was visually inspected for cracks, misassembles or damages to the components.Visual inspection confirmed that the tubing was completely separated from the tubing adapter.Closer inspection of the separation under a lab microscope observed that the adapter had insufficient solvent applied at the engagement during the manufacturing process.The tubing adapter was measured and found to be within specification(s).Functional testing was not necessary as it is obvious that a leak would occur as a result of the separation.The root cause of the separation was a manufacturing issue for insufficient solvent being applied at the engagement of the tubing due to equipment and/ or operator error.
|
|
Event Description
|
It was reported when the rn arrived at the patient's bedside it was observed that the iv that was connected to the patient was broken in two.It was found to be broken about (4) inches below the pump with the other half still connected to the patient.The patient at this time was experiencing confusion, acutely disoriented, anxious, and intervening with other forms of treatment such as picking at dressings covering wounds and pulling out the second iv that was not in use.The rn heard the iv pole being jostled prior to entering the room.The patient grabbed the iv tubing and pulled on it hard enough to break it and the medication (heparin) leaked on the floor.Although it was reported that patient care was delayed, it was confirmed during follow up that this event did not contribute to, or result in any serious adverse impact, nor require any medical intervention for the patient.
|
|
Search Alerts/Recalls
|
|
|