MAUDE Adverse Event Report: BECTON DICKINSON UNSPECIFIED BD IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON UNSPECIFIED BD IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number UNKNOWN
Device Problem Leak/Splash (1354)
Patient Problems Exposure to Body Fluids (1745); Injury (2348)
Event Date 10/21/2019
Event Type  Injury  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.There are multiple bd locations where this unspecified bd device may have been manufactured.A catalog or a lot number could not be confirmed for this incident, and without this information we are unable to determine where the device was manufactured.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.Device expiration date: unknown.Device manufacture date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that a serious injury occurred with an unspecified bd iv catheter.The following information was provided by the initial reporter, "material no.Unknown.Batch no.Unknown.It was reported 2 nurses were cut on the plastic part of the catheter.There is a very small piece of plastic that is left exposed when the catheter full retracts then sticks out just slightly.When it does this, it brings with it a little of the patients blood.On friday, once of my ace iv start personnel was cut by the plastic piece that had some of the patients blood on it.I was told that a similar thing happened to an l&d nurse as well.You can see from a very close inspection of the pic that tiny plastic piece that sticks out.Hello just an update.She (mother-to-be) was stuck by the plastic piece which jets out on top of the iv cath.I am taping it to a piece of paper and sticking it in your mailbox.She was using the cath correctly, no variances.Unfortunately, it did draw blood and she had to go to emp health.During her visit, she was told by emp health staff that someone else from the floor had also been stuck in the same fashion.So 2 rns stuck by these iv caths!!! basically, the plastic retraction sheath retracts so far that a tiny, very sharp piece is left sticking up.Perhaps this is information that can be shared? at this time, there has been no iv reps seen or heard from." 1 of 2 complaints.
 
Event Description
It was reported that a serious injury occurred with an unspecified bd iv catheter.The following information was provided by the initial reporter, "material no.Unknown; batch no.Unknown.It was reported 2 nurses were cut on the plastic part of the catheter.There is a very small piece of plastic that is left exposed when the catheter full retracts then sticks out just slightly.When it does this, it brings with it a little of the patients blood.On friday, once of my ace iv start personnel
 
Manufacturer Narrative
H.6.Investigation summary: one photo was submitted for evaluation.The photo displayed a safety barrel assembly with the needle retracted inside.The unit was taped to a yellow piece of paper, with a black arrow pointing to the top of the grip.The review of the dhr could not be performed as the lot number was not provided.The photo displayed the black arrow pointing to the flange on the white button.The flange is a part of the button and does slight stick up.¿from the customer verbatim : basically, the plastic retraction sheath retracts so far that a tiny, very sharp piece is left sticking up.¿ this is a standard outcome when the needle is retracted.That is in the correct position for the flange on the white button.Conclusion(s): based on the evaluation of the submitted photo; the defect catheter defective/damage could not be refuted nor confirmed.H3 other text : see section h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNSPECIFIED BD IV CATHETER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
MDR Report Key9286949
MDR Text Key171979880
Report Number2243072-2019-02487
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Date Manufacturer Received10/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-