Model Number U125 |
Device Problems
High impedance (1291); Over-Sensing (1438); Device Sensing Problem (2917); High Capture Threshold (3266)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/07/2007 |
Event Type
malfunction
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited high, out-of-range pacing impedance measurements on the left ventricular (lv) channel.This had been occurring since implant; it was noted that the patient was in complete heart block.Lead thresholds were also reportedly high for the past month.A boston scientific technical services (ts) reviewed remote patient monitoring data and confirmed impedance measurements had been jumping to out of range for the past year.No adverse patient effects were reported.This device remains in service.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no objective evidence of a malfunction or inadequate lead-to-device connection; please refer to the description submitted initially for more information regarding the specific circumstances of this event.
|
|
Event Description
|
This report is being filed to submit the results of engineering review of the reported clinical observations.
|
|