Lot Number 0024235175 |
Device Problems
Difficult to Remove (1528); Defective Device (2588)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/16/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device is a combination product.
|
|
Event Description
|
It was reported that difficulty removing a stent device and a tear occurred.The target lesion was located in the moderately calcified and mildly tortuous ostial right coronary artery (rca).A 4.00 x 12mm synergy megatron stent balloon was advanced and successfully deployed in the ostial rca.A stent balloon was used to post dilate the stent at 24 bar.Post dilatation, the stent delivery system (sds) was pulled into the guide catheter.Once the distal part of the sds was completely in the 6f guide catheter, resistance was felt.In response to the resistance, the device was pulled on harder which resulted in a rip somewhere on the sds.The difficulty removing and rip occurred within the guide catheter and therefore the complete system (sds and guide) were safely taken out of the patient.It was noted that the reason for the difficulty removing the device and balloon was likely due to dilating the balloon at 24 bar.No patient complications resulted from this event and the patient was reported to be stable following the procedure.
|
|
Manufacturer Narrative
|
Device is a combination product.Device evaluated by mfr: synergy megatron mr ous 4.00 x 12mm stent delivery system (sds) was returned for analysis, broken into 2 sections and without a stent.An examination found that the balloon appeared to have been subjected to positive and negative pressure.Some bunching of balloon material at proximal end.No visible issues with proximal bond.A visual via scope and tactile examination of the shaft polymer was carried out.A break was identified in the mid shaft extrusion 122mm distal of distal end of strain relief (on the proximal side of the port bond).Stretching of the mid shaft was noted on both sides of the break site.Some stretching also noted on the distal side of the port bond (distal shaft).The break exposed the corewire and separated the distal section.Damage was noted to the distal inner (blue section), it was bunched/accordioned, from 2 to 4mm distal of the bicomponent bond.A visual via scope examination of the tip identified tip damage.A visual and tactile examination of the hypotube found multiple hypotube kinks.A visual via scope examination of the manifold and strain relief did not identify any visible issues.The proximal end of the hypotube was looked down into via scope an no visible blockages were evident.An attempt was made to load the distal section over a 0.014 inch guidewire via tip, the wire was blocked at tip by what appeared to be dried media in the tip.The distal section was placed in small zip lock bag and placed in water batch to soak at 37 degrees to soften dried media.The device was removed from water bath after soaking overnight and second attempt was made to load the wire.The distal section of device was loaded and tracked over wire with some resistance noted, a white substance, appeared to be contract media came out of the wire lumen when tracking the wire.Wire was loaded via tip and exited at the wire exchange port.No other issues were noted.
|
|
Event Description
|
It was reported that difficulty removing a stent device and a tear occurred.The target lesion was located in the moderately calcified and mildly tortuous ostial right coronary artery (rca).A 4.00 x 12mm synergy megatron stent balloon was advanced and successfully deployed in the ostial rca.A stent balloon was used to post dilate the stent at 24 bar.Post dilatation, the stent delivery system (sds) was pulled into the guide catheter.Once the distal part of the sds was completely in the 6f guide catheter, resistance was felt.In response to the resistance, the device was pulled on harder which resulted in a rip somewhere on the sds.The difficulty removing and rip occurred within the guide catheter and therefore the complete system (sds and guide) were safely taken out of the patient.It was noted that the reason for the difficulty removing the device and balloon was likely due to dilating the balloon at 24 bar.No patient complications resulted from this event and the patient was reported to be stable following the procedure.
|
|
Search Alerts/Recalls
|