ABBOTT VASCULAR ARMADA 18 PTA CATHETER PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number 1013466-200
Device Problems Failure to Fold (1255); Difficult to Remove (1528); Material Separation (1562); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/16/2019
Event Type  malfunction  
Manufacturer Narrative
Exemption number e2019001. (b)(4). The device is expected to be returned for evaluation. It has not yet been received. A follow up report will be submitted with all additional relevant information.
 
Event Description
It was reported the procedure was to treat a chronic totally occluded (cto), heavily calcified lesion in the superficial femoral artery. The armada 18 percutaneous transluminal angioplasty (pta) catheter was advanced without resistance noted. The balloon was inflated to 18-22 atmospheres several times and pulled negative twice for deflation. When attempted to remove, there was significant resistance. It appeared the balloon detached from the catheter and was getting stuck in the anatomy. The sheath, guide wire and balloon catheter were all removed as a single unit and the detached portion of the balloon was caught in the sheath. Nothing remained in the patient. A comment was made that the armada 18 balloon doesn't seem to re-wrap well. There were no adverse patient effects and there was no reported clinically significant delay in the procedure. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARMADA 18 PTA CATHETER
Type of DevicePERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key9293032
MDR Text Key176616110
Report Number2024168-2019-13342
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/23/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/30/2021
Device Model Number1013466-200
Device Catalogue Number1013466-200
Device Lot Number8010841
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/08/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/08/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-