Catalog Number 72204354
Device Problem Smoking (1585)
Patient Problem No Patient Involvement (2645)
Event Date 10/18/2019
Event Type  malfunction  
Event Description
It was reported that smoke was coming out of the lens integrated system.Incident occurred before the procedure; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Event Description
Hold by ema 1/3/2020.
 
Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.Complaint of burnt odor was confirmed.Cause of burnt odor is a burnt capacitor on the dsp pcb.Product passed functional testing.The complaint was confirmed and the root cause has been determined to be a burnt capacitor on the dsp pcb.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENS INTEGRATED SYSTEM WIFI VERSION
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key9302693
MDR Text Key165795570
Report Number1643264-2019-00767
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00885556617878
UDI-Public00885556617878
Combination Product (y/n)N
PMA/PMN Number
K153606
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 01/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72204354
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2019
Date Manufacturer Received01/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-