This event has been recorded by zimmer biomet under (b)(4).Udi no: (b)(4).The device history record (dhr) for intellicart system serial number (b)(4) was reviewed and noted no related non-conformances, requests for deviation (rfd), change notices (cn), or any other issues with manufacturing.The dhr review found no issues with the device and all verifications, inspections, and tests were successfully completed.On 08 november 2019, a zimmer biomet sales rep requested that the cart be exchanged due to the complaint; the cart was not noted to be evaluated prior to the exchange request.An exchange order was subsequently written and submitted.A new cart (serial number (b)(4)) was shipped to the account and arrived on 12 november 2019.A zimmer biomet authorized service technician was dispatched to the account to complete the exchange, and confirmed the exchange was completed with the new unit in service without issue on 14 november 2019.The old cart was shipped back to riverside and arrived on 20 november 2019.Crm notes that the cart was exchanged by request of the rsm and sales rep; the odor concern was never confirmed.The unit will be refurbished with new serial number (b)(4); no additional information is available at this time.Because the reported event was never confirmed as a technician did not evaluate the unit for the odor concern and the unit was exchanged simply by request, it is not known what caused the reported odor issue.Therefore, a specific root cause cannot be determined.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.Based on the information provided, this investigation determined that there is no need for further action at this time, as this complaint is determined to not be a new confirmed quality or manufacturing issue.This complaint will be tracked and trended for any adverse trends that may require additional actions.
|