SMITH & NEPHEW, INC. R3 20DEG 36MM TRIAL56MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Catalog Number 71367956 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/29/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that during a thr procedure, the sharp edges of the trial cut the gloves of the surgeon.Plastic liner had gouges.No delay.Backup device available.No impact or injury reported.
|
|
Manufacturer Narrative
|
The associated complaint device was returned and evaluated.A visual inspection confirms the trial has multiple scratches, burrs and deep gouges in the plastic.The device shows extreme signs of wear/usage.There is no lot number available for this device.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.
|
|