The device has been received at coloplast; however the evaluation is not yet complete.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Once our evaluation is complete, a follow-up report will be submitted.
|
This follow-up mdr is created to document the evaluation of the returned device and corrected implant date.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.A titan touch pump and two cylinders were received for evaluation.The inlet tube with connector was detached for testing.Examination and testing of the returned components revealed a separation within abrasion on the inlet tube of the pump.Testing revealed this to be a site of leakage.Partial separations within abrasion were noted on both exhaust tubes of the pump.Testing revealed these to not be sites of leakage.Abrasion was also noted on the exhaust tube of cylinder 1.No functional abnormalities were noted with either cylinders or detached inlet tube with connector.Based on examination of the returned product, it was concluded that while in-vivo all pump tubing had overlapped and abraded against one another.This positioning, in combination with device usage over time, could contribute to sufficient stress(s) to separate the inlet tubing of the pump.A separation of this type would then allow the loss of fluid, making the device inoperable.
|