RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Defective Device (2588)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 15nov2019.
 
Event Description
The customer reported unit has a hi ground reading.There was no patient involvement.The manufacturer's field service engineer (fse) confirmed the reported issue.The fse replaced the defective ac power cord to address the reported problem.The unit successfully passed the required performance verification test.
 
Manufacturer Narrative
G4: 09mar2020 b4: 1(b)(6)2020.The ac power cord was returned to failure investigation (fi) for evaluation.The visual inspection of the ac power cord revealed cord has braided appearance and bent contacts on plug end.The determination could be made that the device failed to meet specifications.The device was not being used for treatment when the reported event occurred, and there is a relationship of the device to the reported problem.Ac power cord failed 4-wire resistance test.Verified reported failure.High pin-pin resistances noted g-g.Root cause determined to be wear and tear.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key9329650
MDR Text Key190361313
Report Number2031642-2019-10541
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 11/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2020
Was the Report Sent to FDA? No
Date Manufacturer Received11/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-