Model Number 232447
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/28/2019
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This is report 2 of 2 for the same event.It was reported by the affiliate in (b)(6) that during posterior cruciate ligament (pcl) surgical procedure, it was observed that while fixing the tibial end, the plate could not be tightened.According to the report, the loop was loose.It was reported that both devices had to be removed.It was not reported if there was a delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: correction: the date device returned to manufacturer was documented as (b)(6) 2019 on the initial report and has been updated as (b)(6) 2019.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: according to the information provided, it was reported that during during the pcl reconstruction, when fix the tibial end, the plate could not be tightened, the loop was loose.The complaint device was received and evaluated.Visual observation was performed, can¿t confirm that the suture could not be tightened because the suture was not threaded through the holes in the implant.  a manufacturing record evaluation was performed for the finished device lot/serial number (b)(6), and no non-conformances were identified.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RGDLOOP ADJUSTABLE STND
Type of Device
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9348222
MDR Text Key207938280
Report Number1221934-2019-59620
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705024094
UDI-Public10886705024094
Combination Product (y/n)N
PMA/PMN Number
K140324
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup,Followup
Report Date 10/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2022
Device Model Number232447
Device Catalogue Number232447
Device Lot Number5L45073
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2019
Date Manufacturer Received01/27/2020
Is This a Reprocessed and Reused Single-Use Device? No