MEDOS INTERNATIONAL SÃ RL RGDLOOP ADJUSTABLE STND; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
|
Back to Search Results |
|
Model Number 232447 |
Device Problem
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/28/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
This is report 2 of 2 for the same event.It was reported by the affiliate in (b)(6) that during posterior cruciate ligament (pcl) surgical procedure, it was observed that while fixing the tibial end, the plate could not be tightened.According to the report, the loop was loose.It was reported that both devices had to be removed.It was not reported if there was a delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: correction: the date device returned to manufacturer was documented as (b)(6) 2019 on the initial report and has been updated as (b)(6) 2019.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: according to the information provided, it was reported that during during the pcl reconstruction, when fix the tibial end, the plate could not be tightened, the loop was loose.The complaint device was received and evaluated.Visual observation was performed, can¿t confirm that the suture could not be tightened because the suture was not threaded through the holes in the implant. a manufacturing record evaluation was performed for the finished device lot/serial number (b)(6), and no non-conformances were identified.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
|
|