Catalog Number 51006004L
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/31/2019
Event Type  malfunction  
Manufacturer Narrative
A review of the manufacturing documentation associated with lot 17628256 presented no issues during the manufacturing process that can be related to the reported event. Additional information is pending and will be submitted within 30 days upon receipt.
 
Event Description
As reported, the 6mm 4cm saber rapid exchange (rx) percutaneous transluminal angioplasty (pta) balloon catheter (bc) was used for pre-dilation but there were anomalies confirmed at the mounting position of the balloon marker. It was confirmed that the distal and proximal marker were out of position on angiography. There was no reported patient injury. The lesion was at the left superficial femoral artery (sfa). A contralateral approach was made from the right femoral artery. The guidewire crossed the lesion and the device was inflated to eight atmospheres (8atm). The procedure was completed using a same size non-cordis balloon catheter, which was used for a post dilation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSABER RX6MM4CM155
Type of DeviceCATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, co. tipperary
EI
Manufacturer (Section G)
CORDIS CASHEL
cahir road
cashel, co. tipperary
EI
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014
7863138372
MDR Report Key9350757
MDR Text Key198970390
Report Number9616099-2019-03365
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 01/02/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date11/30/2019
Device Catalogue Number51006004L
Device Lot Number17628256
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/20/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-