SMITH & NEPHEW, INC. UNKNOWN SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number UNKNOWN
Device Problem Overheating of Device (1437)
Patient Problem Burn(s) (1757)
Event Date 09/24/2019
Event Type  Injury  
Event Description
It was reported that during a left knee arthroscopy with arthroscopic irrigation and debridement, an odor was noted by the nurse. Operating room staff noted the shaver handle to be hot and smoke coming from the shaver cord. The tower was disconnected from the main wall receptacle. Surgeon continue closing, with not significant delay. Post-operatively patient was examined and a red raised area was noted on her right knee. Surgeon applied silvadene ointment and dry dressing to the affected area. Patient was seen once by vascular surgeon as an outpatient and burn was nearly healed up. Recommendation to continue current treatment of silvadene. All available information has been disclosed. If additional information should become available, a supplemental report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key9350823
MDR Text Key167314859
Report Number1643264-2019-00815
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date 01/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/23/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

Patient Treatment Data
Date Received: 11/20/2019 Patient Sequence Number: 1
-
-