The investigation determined that higher than expected vitros glucose (glu) results were obtained from vitros performance verifiers (pv) processed using two different lots of vitros chemistry products glu slides on a vitros 5600 integrated system.The most likely cause of the higher than expected results is suboptimal calibrations likely caused by the customer not following ortho recommended protocol for preparation of the vitros calibrator kit 1 lot 0168.The customer likely used a calibrator that was past stability limits to calibrate the vitros glu reagent lots which led to the attainment of higher than expected quality control (qc) results.E-connectivity data also showed the customer was found to be calibrating too often even when qc was well within acceptable range based on historical quality control results, a vitros glu lot 0049-0967-1822 performance issue is not a likely contributor to the event as qc results showed the reagent to be accurate.The customer exhausted their inventory of vitros glu lot 0049-0967-1822 before a precision test could be run.There were no historical quality results for vitros glu lot 0049-0967-1888 because it was a new reagent lot and had only recently been calibrated and put into use.There is no evidence that the vitros 5600 integrated system malfunctioned.
|