Model Number MOSAIQ
Device Problem Computer Software Problem (1112)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Event Description
The customer reported that the treatment record was not sent to (b)(6).
 
Manufacturer Narrative
The investigation was completed by conducting a thorough evaluation of the product and the reported information.The investigation found that by clicking "chart/capture" to open the treatment chart, but then closing the treatment chart without recording a dose will still create a snapshot in table dosesummarysnapshot.This allows for duplicate dosesummarysnapshots which can confuse the exporting charge information.Mosaiq contains the correct information regarding how a patient has been treated and will be treated going forward.This issue would not result in a mistreatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Type of Device
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA INC
100 mathilda place
5th floor
sunnyvale CA 94086
MDR Report Key9353856
MDR Text Key188493014
Report Number2950347-2019-00021
Device Sequence Number1
Product Code IYE
UDI-Device Identifier00858164002091
UDI-Public00858164002091
Combination Product (y/n)N
PMA/PMN Number
K141572
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 01/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2019
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMOSAIQ
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-