Model Number 60-02-60
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/24/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement. Livanova (b)(4) manufactures the x. The incident occurred in x. A livanova field service representative was dispatched to the facility to investigate the device and he could not confirm the reported event. The device was working according to the specifications. The technician checked the operation record and connect data and found that, when the event occurred, the speed was at the minimum speed (1000 rpm ) and not zero. However, the device was replaced with another as precaution. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that the speed of a centrifugal pump 5 (cp5) became zero during procedure and no alarm was displayed. The flow rate dropped to zero triggering backflow alarms. The user turned the knob to restore the flow. After a while, the issue recurred and the flow rate was restored by turning the knob again. The procedure could be completed without further issues. There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCENTRIFUGAL PUMP 5 (CP5)
Type of DeviceCONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9355367
MDR Text Key202999005
Report Number9611109-2019-00907
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/21/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number60-02-60
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/10/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/19/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

-
-