• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1601-05
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Event Description
The recipient reportedly experienced lack of benefit from the device.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device on the contralateral side.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient has reportedly healed.The external visual inspection revealed silicone damage on the top cover, slices near the fantail and electrode ground ring, and the electrode was severed near the array prior to receipt.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey mcdonald
28515 westinghouse place
valencia, CA 91355
MDR Report Key9355803
MDR Text Key167407119
Report Number3006556115-2019-00702
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016862267
UDI-Public(01)07630016862267(11)190411(17)220331
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date03/31/2022
Device Model NumberCI-1601-05
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/12/2019
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/11/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-