It was reported that on morning of (b)(6) 2019 during system control, it was noticed that very bloody condensation water emerged from the gas outlet of the oxygenator (about 20 ml).Careful blowing of the oxygenator at 7l / min resulted in further bloody water.That's why the hls set was changed.The defective kit (lot 70132912, be-015703112) was then rinsed with tap water and ready for collection and further assessment by getinge.A patient's risk of the defect is extremely unlikely, since the change took place uncomplicated within a few minutes.Internal ref.No.(b)(4), onesupport: (b)(4).
|