TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II) CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING Back to Search Results
Model Number 4415
Device Problem Device Alarm System (1012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/28/2019
Event Type  Malfunction  
Manufacturer Narrative

The field service representative (fsr) verified the reported issue. He replaced the pressure switch which resolved the issue. The unit operated to the manufacturer's specifications. The suspect part was returned to the manufacturer for further evaluation.

 
Event Description

It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the heater cooler would not stop alarming. As a result, an alternate device was employed. The surgical procedure was completed successfully. There was no delay, no blood loss, nor adverse consequences to the patient.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key9362753
MDR Text Key199110587
Report Number1828100-2019-00619
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial,Followup
Report Date 03/17/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received11/22/2019
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number4415
Device Catalogue Number4415
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer11/04/2019
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received02/27/2020
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured04/26/2007
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-