The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, the oxygenator was working but flow reduced to about 1.
5 l per minute.
As per the user facility, they received an email about the fx15 being swapped out due to a pressure excursion.
As per the perfusionist, the patient had an adequate anti coagulation and good act xclamp still on about 20 minutes into the case.
The patient was warm, and believed that the changes rose over a short period of 300 with the post pressure rising to around 500 ml hg.
They were able to swap out the fiber bundle with a 30 ml blood loss as they sequestered almost all of it into the reservoir.
The oxygenator was replaced and it performed well without pressure excursion, and the case was continued without further incident or patient harm.
No known impact or consequence to patient.
There was a delay for approximately 10 minutes while changing out.
The procedure was completed successfully.
|