Model Number UNK_NAV_SYS |
Device Problem
Output Problem (3005)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/04/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product: sw version: 1.2.0.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system.It was reported outside of a procedure that the next task button would sometimes temporarily appear grayed/disabled after transitioning to the acquire images task.It was noted that the site transitioned to acquire images task with a previous exam downloaded on the system, monitored the next task button (navigate in this case), and the button would flicker between disabled (grey) and enabled (blue) before stabilizing on blue.There was no patient involvement.
|
|
Manufacturer Narrative
|
The software investigation found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|