• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC ENSITE VELOCITY SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC ENSITE VELOCITY SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 100014514
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report further information regarding the event were requested but not received.
 
Event Description
Prior to the procedure, the amplifier started the self-test at start-up, but it was not completed successfully.The red light was lit permanently or flashing.The amplifier was power cycled, disconnected from the power, and the cables were reconnected, however resolution was not found., the problem could not be solved.The patient had been prepped and the procedure was cancelled.There were no adverse patient consequences.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported startup issue and subsequent cancellation could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY SYSTEM VELOCITY AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC
one st. jude medical drive
st. paul MN 55117
MDR Report Key9382948
MDR Text Key168214109
Report Number2184149-2019-00236
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05414734210713
UDI-Public05414734210713
Combination Product (y/n)N
PMA/PMN Number
K141050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number100014514
Device Catalogue Number100014514
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-