• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) RIATA ACTIVE FIXATION; DEFIBRILLATION LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) RIATA ACTIVE FIXATION; DEFIBRILLATION LEAD Back to Search Results
Model Number 1580/65
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/18/2019
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer reference number: 2938836-2019-16782.It was reported that an asymptomatic patient presented for a generator change out procedure.During procedure, some erosion was noted on the right atrial lead, and externalized conductor cables were noted on the right ventricular lead.The sensing, threshold and impedance measurements were stable on both the leads.The lead issues were confirmed via chest x-ray.The right atrial lead was repaired using medical adhesive and a lead end cap.No intervention was performed on the right ventricular lead.Both the leads were connected to a new device.The patient was stable before, during and post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIATA ACTIVE FIXATION
Type of Device
DEFIBRILLATION LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9392639
MDR Text Key168599027
Report Number2938836-2019-16781
Device Sequence Number1
Product Code NVY
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date 11/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2007
Device Model Number1580/65
Device Lot Number0002164554
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/20/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/19/2004
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0457-2012
Patient Sequence Number1
Patient Age56 YR
Patient Weight62
-
-