Model Number 661440 |
Device Problems
Break (1069); Material Twisted/Bent (2981); Device Handling Problem (3265)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
|
Event Date 11/05/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device is not available for return, device is still implanted.
|
|
Event Description
|
Primary left tibia procedure.It was reported that the 3.5 cortical screw was being inserted outside of the 20° cone of angulation in the shaft and was visibly bending during insertion.It is not known if the screw was aligned with the drill hole in the far cortex.The screw broke inside the patient.As the head was engaged with the plate and was not sitting proud, the surgeon made the decision to leave the entirety of the screw in the patient.Surgery was completed successfully with no delay.The rep may be able to obtain x-rays, and otherwise confirmed that no further information will be released by the hospital or surgeon.
|
|
Manufacturer Narrative
|
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.The device history record could not be reviewed because the affected lot number was not communicated.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
|
Event Description
|
Primary left tibia procedure.It was reported that the 3.5 cortical screw was being inserted outside of the 20° cone of angulation in the shaft and was visibly bending during insertion.It is not known if the screw was aligned with the drill hole in the far cortex.The screw broke inside the patient.As the head was engaged with the plate and was not sitting proud, the surgeon made the decision to leave the entirety of the screw in the patient.Surgery was completed successfully with no delay.The rep may be able to obtain x-rays, and otherwise confirmed that no further information will be released by the hospital or surgeon.
|
|
Search Alerts/Recalls
|