Model Number 209063 |
Device Problems
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/08/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
As part of the normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
|
Event Description
|
Handle fell off during use.Case type: tka.
|
|
Event Description
|
Handle fell off during use.Case type: tka.
|
|
Manufacturer Narrative
|
Reported issue: handle fell off during use.Case type: tka.Product inspection: visual inspection.Mics-209063 sn#(b)(6); rma#(b)(4).Inspected per d06917 and determined failure of the following test step.Sec# 7.1.2.Visual missing handle.Disposition: rtv.Inspected by: (b)(6).Device history review: review of device history record indicate that (b)(4) devices were manufactured under lot 42020319 and (b)(4) accepted and (b)(4) rejected on 04/04/2019.Review of (b)(4) revealed that the non conformance is not related to the failure alleged in this complaint.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 209063, lot number42020319 shows no additional complaints related to the failure in this investigation.Conclusion: the alleged failure mode was confirmed.No additional investigation or specific actions are required.¿ if additional information is received then the complaint will be reopened.Nc/capa review: a review of stryker¿s nc/capa database indicated there have been an nc and capa associated with the product and failure mode reported in this event.This is (b)(4) and capa 1452931.
|
|
Search Alerts/Recalls
|