Model Number 24691
Device Problems Material Rupture (1546); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/15/2019
Event Type  malfunction  
Event Description
It was reported that balloon rupture occurred the 100% stenosed target lesion was located in the moderately tortuous and severely calcified below the knee.A 1.5mm x 20mm x 143cm coyote es balloon catheter was advanced for dilation.However, during third inflation at 8 atmospheres, the balloon ruptured.The procedure was completed with another device.There were no patient complications reported.
 
Event Description
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the moderately tortuous and severely calcified below the knee.A 1.5mm x 20mm x 143cm coyote es balloon catheter was advanced for dilation.However, during third inflation at 8 atmospheres, the balloon ruptured.The procedure was completed with another device.There were no patient complications reported.
 
Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a coyote es balloon catheter.There were numerous kinks throughout the catheter.Microscopic inspection revealed a tear over the markerband.The torn material was pulled up which indicates that the balloon became caught on either calcification or another device during withdrawal.The reported balloon rupture was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE ES
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9410524
MDR Text Key175082184
Report Number2134265-2019-14908
Device Sequence Number1
Product Code LIT
UDI-Device Identifier08714729767169
UDI-Public08714729767169
Combination Product (y/n)N
PMA/PMN Number
K093636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 01/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24691
Device Catalogue Number24691
Device Lot Number0023485730
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2019
Date Manufacturer Received12/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-