Model Number 24691 |
Device Problems
Material Rupture (1546); Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/15/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that balloon rupture occurred the 100% stenosed target lesion was located in the moderately tortuous and severely calcified below the knee.A 1.5mm x 20mm x 143cm coyote es balloon catheter was advanced for dilation.However, during third inflation at 8 atmospheres, the balloon ruptured.The procedure was completed with another device.There were no patient complications reported.
|
|
Event Description
|
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the moderately tortuous and severely calcified below the knee.A 1.5mm x 20mm x 143cm coyote es balloon catheter was advanced for dilation.However, during third inflation at 8 atmospheres, the balloon ruptured.The procedure was completed with another device.There were no patient complications reported.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: returned product consisted of a coyote es balloon catheter.There were numerous kinks throughout the catheter.Microscopic inspection revealed a tear over the markerband.The torn material was pulled up which indicates that the balloon became caught on either calcification or another device during withdrawal.The reported balloon rupture was confirmed.
|
|