• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTICSUN 5000 ARCTIC SUN DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTICSUN 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000
Device Problems Failure to Sense (1559); Device Sensing Problem (2917)
Patient Problems Therapeutic Effects, Unexpected (2099); Alteration In Body Temperature (2682); Patient Problem/Medical Problem (2688)
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete, a supplemental report will be filed. The device was not returned.
 
Event Description
It was reported that the device had a sensing issue during the warming process. The nurse stated the patient was rewarming too fast. There was good flow, and a good fit to pads. The patient's temperature was 35. 9c, rewarm was reading from 34c at 0. 25c/hr. , the water temperature was 6c and the trend indicator was one arrow up. The patient was actively shivering, neuromuscular blocker was started but the patient was still shivering. Ms&s explained that the device wasn't rewarming too fast and the patient's temperature was higher than it should be because of shivering. Per sicu follow up, the patient was able to complete therapy and the device was sent to biomed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTICSUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key9410698
MDR Text Key169096595
Report Number1018233-2019-07688
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741080142
UDI-Public(01)00801741080142
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 12/04/2019 Patient Sequence Number: 1
-
-