Model Number 284004 |
Device Problem
Suction Failure (4039)
|
Patient Problem
Not Applicable (3189)
|
Event Date 11/22/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
|
|
Event Description
|
It was reported by the affiliate via phone that during a shoulder arthroscopic the fms vue pump-shaver box didn¿t suction properly.No patient consequence, however there was a surgical delay of 10 minutes.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary = > the device was received and evaluated at the service center.Neither the fault reported by the customer could be verified nor another fault was found with the device upon evaluation.However, it was found that the tamper-proof seal was missing.The device was cleaned, tested and found to be fully functional.Since the reported condition is not confirmed,the root cause for the reported failure cannot be determined.The missing tamper-proof seal is an indication that someone not authorized has opened the device.A manufacturing record evaluation was performed for the finished device (serial number : (b)(6)), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot = > a manufacturing record evaluation was performed for the finished device (serial number : (b)(6)), and no non-conformances were identified.
|
|