Model Number CYF-5
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/30/2019
Event Type  Malfunction  
Manufacturer Narrative

The subject device in this report has not been returned to omsc for evaluation. The exact cause of the reported event could not be conclusively determined at this time. If additional information becomes available, this report will be supplemented.

 
Event Description

Olympus medical systems corp. (omsc) was informed that as a result of multiple microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device. [first time; (b)(6) 2019]. Instrument channel: unspecified microbes (1 cfu). [second time; (b)(6) 2019]. Distal end: unspecified microbes (1 cfu) [third time; (b)(6) 2019]. Instrument channel: unspecified microbes (2 cfu). [fourth time; (b)(6) 2019]. Instrument channel: unspecified microbes (2 cfu). The device had been reprocessed with a non-olympus automated endoscope reprocessor, soluscope, using peracetic acid. There was no report of infection associated with this report.

 
Manufacturer Narrative

This supplemental report is being submitted to provide additional information. The subject device had been reprocessed with a non-olympus automated endoscope reprocessor, soluscope serie 4, using peracetic acid. The exact cause of the reported event could not be conclusively determined at this time. If additional information becomes available, this report will be supplemented.

 
Manufacturer Narrative

This supplemental report is being submitted to provide additional information. The subject device has not been returned to omsc but was returned to olympus (b)(4). (b)(4) sent the device to a third party laboratory for microbiological testing. As a result of the testing, no microbe was detected from the sample collected from all channels of the device. The testing result cleared the (b)(6) guideline. The exact cause of the reported event could not be conclusively determined at this time. If additional information becomes available, this report will be supplemented.

 
Manufacturer Narrative

This supplemental report is being submitted to provide additional information. Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity. The exact cause of the reported event could not be conclusively determined. If additional information becomes available, this report will be supplemented.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFIBERSCOPE "CYF-5", EUROPEAN VERSION
Type of DeviceCYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key9420518
MDR Text Key219787576
Report Number8010047-2019-04170
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeFR
PMA/PMN NumberK032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type FOREIGN,HEALTH PROFESSIONAL,U
Reporter Occupation
Type of Report Initial,Followup,Followup,Followup
Report Date 04/14/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received12/05/2019
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberCYF-5
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received03/19/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured03/26/2014
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-