ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Device Problems Break (1069); Unintended Ejection (1234); Physical Resistance/Sticking (4012)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 11/21/2019
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: complaint history and product history records were reviewed and documentation indicated the product met release criteria. Root cause has not been identified. There have been no other complaints reported in the lot number. Additional information has been requested. The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported during the implant of an intraocular lens (iol) using a preloaded delivery system, the surgeon felt a bit of resistance when pushing the plunger. The lens then shot out and the rear haptic was broken and left inside the patient's anterior chamber. There is no reported patient harm. The lens was removed and replaced with a backup lens. Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Type of DeviceLENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jason michaelides
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175686438
MDR Report Key9426744
MDR Text Key176113417
Report Number1119421-2019-02075
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/12/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date02/28/2021
Device Model NumberAU00T0
Device Lot Number12605352
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/22/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial

-
-