Catalog Number 283512
Device Problem Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the device was received and evaluated at the service center.The reported complaint that the device does not turn was confirmed.It was found that there was a short circuit in the motor cable and the insulation resistance of the motor cable was out of tolerance.The motor cable was replaced and the device was tested and found to be working according to specifications.However, given the information provided we cannot discern a definitive root cause for the short circuit and damage to the motor cable.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on (b)(6) 2018 and passed all functional testing before being returned to the customer.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Event Description
It was reported by the affiliate via phone that during an unknown procedure the motor of the micro tornado hp w handcontrol was jammed and it didn¿t turned.The procedure was completed using a replacement.No patient consequence and no surgical delay was reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key9427864
MDR Text Key207613993
Report Number1221934-2019-59871
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeLO
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Reporter Occupation Other
Type of Report Initial
Report Date 12/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2019
Date Manufacturer Received12/03/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N