• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON AND COMPANY BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON AND COMPANY BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 306575
Device Problems Loose or Intermittent Connection (1371); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 09/23/2019
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that foreign matter occurred before use with a bd vacutainer® sst¿ ii advance plus blood collection tubes.The following information was provided by the initial reporter, "the cap is crushed and dirty.".
 
Manufacturer Narrative
Investigation summary: one sample and two photos were received for evaluation.The sample has the tip cap bent-damaged and dirty.This occurs when the syringe is misplaced in the sterilizer tray.The next tray that goes on top lays on top of the misplaced syringe tip cap inducing the damage and getting the tip cap dirty.The two photos show the syringe with the tip cap defects.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.
 
Event Description
It was reported that foreign matter occurred before use with a bd vacutainer® sst¿ ii advance plus blood collection tubes.The following information was provided by the initial reporter, "the cap is crushed and dirty.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON DICKINSON AND COMPANY
2153 12th avenue
columbus NE 68601
MDR Report Key9434134
MDR Text Key191436927
Report Number1911916-2019-01289
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
PMA/PMN Number
BK050036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 12/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2022
Device Catalogue Number306575
Device Lot Number9080859
Date Manufacturer Received11/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-