Model Number 2426-0007 |
Device Problems
Material Separation (1562); Component Missing (2306)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/05/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Although requested, the affected product has not been received.A follow up report will be submitted with investigation results should the devices be received for evaluation.
|
|
Event Description
|
It was reported that the tubing is missing blue rings which causes the tubing to separate during use.There was no harm to patient.
|
|
Manufacturer Narrative
|
Correction: d.4.Additional information added; d.10, d.11, & h.6.The customer¿s report that the tubing was missing the blue ring which caused the tubing to separate was confirmed.The set was visually inspected for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components.Visual inspection observed that the silicone segment was separated from the lower fitment at the pump segment.The ring retainer was not received with the set and a ring retainer indentation was not observed around the end of the silicone pump tubing where the separation occurs.Closer inspection under a lab microscope observed no indication of damage or tool marks on the pump segment.Functional testing could not be performed due to the separation and obvious leaking that would occur.The silicone pump tubing and fitment were found to be within measurement specification(s).The root cause of the primary set silicone segment separation was due to the set's retainer ring not assembled onto the set during manufacturing assembly process.
|
|
Event Description
|
It was reported from the pediatric bone marrow unit that the tubing was missing blue ring which causes the tubing to separate during use.There was no harm to patient.
|
|
Manufacturer Narrative
|
Additional information provided in b.3.
|
|
Event Description
|
It was reported from the pediatric bone marrow unit, that the tubing was missing a blue ring, which caused the tubing to separate during use.It was further confirmed during follow up, that there was no patient harm or impact as a result of this event.
|
|
Search Alerts/Recalls
|