Model Number MMT-1715K
Device Problem Obstruction of Flow (2423)
Patient Problems Hyperglycemia (1905); Reaction, Injection Site (2442)
Event Date 12/03/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Customer reported via phone call that they experienced high blood glucose.The customer blood glucose level was 568 mg/dl at the time of incident.Customer reported that the tape did not stick using her infusion set.Customer received insulin flow block during bolus and basal.Customer reported that the insulin pump alarmed with no reservoir detected.Customer reported insulin exits the tubing.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan as per health care professional instruction.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device passed the displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test and self test.No unexpected insulin flow blocked alarm/no delivery alarm noted during testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key9434606
MDR Text Key169807832
Report Number2032227-2019-128781
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00763000190439
UDI-Public(01)00763000190439(17)220603
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date06/03/2022
Device Model NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device Lot NumberHG3GHSN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received01/31/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/05/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age34 YR
Patient Weight277
-
-