MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000E
Device Problems Device Displays Incorrect Message (2591); Device Sensing Problem (2917); Therapeutic or Diagnostic Output Failure (3023)
Patient Problems No Consequences Or Impact To Patient (2199); Inadequate Pain Relief (2388)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the arctic sun device began cooling the patient after being in rewarm. The patient's temperature was 35. 1c with a target temperature of 37c. The water temperature was 21c. The flow rate was stopped, as the nurse had stopped the machine. The event log showed an alert 11 (patient temperature below low patient alert) and alarm 14 (temperature probe out of range). Therapy was restarted and the water temperature began increasing, but backed off for a short time. The flow rate settled at 3. 3l/min. There were two small universal pads in place as well as a bair hugger. Per troubleshooting, the device was placed in manual control, and the water temperature easily increased to 40c. Per reviewing the graph, someone had changed the target temperature to 33c. The normothermia rate was set for 0. 5c/hour, and the nurse stated she would watch and ensure the patient trends up appropriately. The device was placed back in automatic mode and therapy continued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key9444947
MDR Text Key177645997
Report Number1018233-2019-07846
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741127755
UDI-Public(01)00801741127755
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/09/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000E
Device Catalogue Number50000000E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/04/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/08/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-