CORDIS CASHEL TRAPEASE PVCF FEM/JUG 55CM CSI; FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number 466P306AU
Device Problems Failure to Align (2522); Migration (4003)
Patient Problems Vessel Or Plaque, Device Embedded In (1204); Perforation of Vessels (2135)
Event Date 12/10/2016
Event Type  Injury  
Manufacturer Narrative
Occupation: other, senior counsel, litigation.Please note that the exact event date is unknown and the event date is the complaint awareness date.As, the patient underwent placement of a trapease inferior vena cava (ivc) filter.The indication for filter placement is not available.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to perforation of multiple filter struts, migration of the filter, embedded, tilting and unable to be retrieved.The filter remains implanted; thus, unavailable for analysis.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without procedural films for review, the filter tilt reported could not be confirmed.Additionally, the timing and mechanism of the filter tilt is unknown.Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness.It was reported that there was perforation of the ivc; however, a clinical conclusion could not be determined as to the cause of the event.A review of the instructions for use notes vessel damage such as intimal tears and perforation as procedural complications related it ivc filters.Ivc perforation from filters is relatively common, and directly related to how long the filter has been in place.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Inferior vena cava (ivc) filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the instruction for use (ifu) as such.Possible causes for filter migration include mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the inferior vena cava including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.The predominant concern for embedding with in the wall of the ivc is the development of endothelialization.Endothelialization is the healing of the inner surfaces of vessels or grafts by endothelial cells.This is the normal process whereby the body heals and recovers from invasive procedures.Endothelialization has been shown to occur in as short a period as 12 days.Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics.Without procedural films or images for review the reported event(s) could not be confirmed.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.Please note that this is the initial/final report for this product.
 
Event Description
As reported by the legal brief, the patient underwent placement of a trapease vena cava filter.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to perforation of multiple filter struts, migration of the filter, the filter was embedded, tilting of the filter and unable to be retrieved.As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.The form states that the device was unable to be retrieved, but no documentation of any attempt to remove the device was provided.
 
Manufacturer Narrative
After further review of additional information received, the following sections have been updated accordingly: a2, b3, b4, b7, d1, d4, d11, g3, g4, g7, h1, h2, h4 and h6.Section b5: additional information received per the medical records indicate that the patient has a history of multiple trauma with pelvic and long bone fracture.The filter was deployed via right common femoral vein.It was placed near the l2-l3 vertebral body.The patient tolerated the procedure well.  additional information received per the patient profile form (ppf) states that the patient experienced perforation of inferior vena cava (ivc), the filter was embedded, unable to be retrieved ,and pain post implant.These injuries have caused emotional distress, mental anguish, anxiety and stress.The patient became aware of the filter tilt ten years and six months after the index procedure.Although it was reported that the device was unable to be retrieved, there were no documented attempts to remove the device.Additional information is pending and will be submitted within 30 days of receipt.
 
Manufacturer Narrative
After further review of additional information received, the following sections have been updated accordingly: b4, g4, g7, h1, h2 and h6.As reported, the patient underwent placement of a trapease vena cava filter.The patient is reported to have had a history of trauma with multiple pelvic and long bone fractures.The indication for the filter placement was not reported.The filter was implanted via right common femoral vein and placed at the level of the l2-l3 vertebral body.The patient is reported to have tolerated the procedure well.Approximately ten and a half years after the filter implantation, the patient became aware that had tilted, migrated and was embedded.The patient also reported that multiple struts had perforated the inferior vena cava (ivc).In addition, the patient indicated that the filter could not be retrieved; though attempts to retrieve it were not documented.The patient further reported having experienced pain, emotional distress, mental anguish, anxiety and stress associated with the filter.The product was not returned for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the ivc for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without images or procedural films for review, the reported filter tilt, migration, device embedment and perforation events could not be confirmed and the exact cause could not be determined.Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness.Additionally, the timing and mechanism of the filter tilt is unknown.Ivc filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the instruction for use (ifu) as such.Possible causes for filter migration include mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the ivc including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.The predominant concern for embedding within the wall of the ivc is the development of endothelialization.Endothelialization is the healing of the inner surfaces of vessels or grafts by endothelial cells.This is the normal process whereby the body heals and recovers from invasive procedures.The trapease vena cava filter is designed for permanent implantation.Endothelialization has been shown to occur in as short a period as twelve days.It is unknown if the tilt contributed to the reported perforation.A review of the ifu notes vessel damage such as intimal tears and perforation as procedural complications related to ivc filters.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Due to the nature of the complaint, the reported pain experienced by the patient could not be confirmed and the exact cause could not be determined.These clinical events do not represent evidence of a device malfunction.Clinical factors that may have influenced these events include the patient¿s pre-existing co-morbidities, pharmacological issues and lesion characteristics.The anxiety experienced by the patient does not represent a device malfunction.Anxiety, part of the body¿s natural response to stress and can cause feelings of, but not limited to, nervousness, mental anguish, fear, unease and worry.Given the limited information available for review, there is nothing to suggest that a malfunction in the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEASE PVCF FEM/JUG 55CM CSI
Type of Device
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel co. tipperary
EI 
MDR Report Key9445737
MDR Text Key170696509
Report Number1016427-2019-03642
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2008
Device Catalogue Number466P306AU
Device Lot NumberR0106415
Was Device Available for Evaluation? No
Date Manufacturer Received02/14/2020
Patient Sequence Number1
Treatment
UNKNOWN 6 FRENCH SHEATH; UNKNOWN 0.035 GLIDE WIRE
Patient Outcome(s) Life Threatening;
Patient Age47 YR
-
-