Model Number 3662 |
Device Problems
Disconnection (1171); Low impedance (2285)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/25/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
Related manufacturer reference number: 1627487-2019-12837.Related manufacturer reference number: 1627487-2019-12838.It was reported the patient's ipg was unable to enter mri mode.Diagnostic testing revealed several low impedance values.In tun, surgical intervention was undertaken on (b)(6) 2019 wherein during intra-op testing it was observed that one of the leads was not fully inserted into the ipg header.Physician was able to clean off the lead and reinsert.Mri mode was turned on.This addressed the issue.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not accessible for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Related manufacturer reference number: 1627487-2019-12837.Related manufacturer reference number: 1627487-2019-12838.
|
|