Model Number 3662
Device Problems Disconnection (1171); Low impedance (2285)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/25/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Related manufacturer reference number: 1627487-2019-12837.Related manufacturer reference number: 1627487-2019-12838.It was reported the patient's ipg was unable to enter mri mode.Diagnostic testing revealed several low impedance values.In tun, surgical intervention was undertaken on (b)(6) 2019 wherein during intra-op testing it was observed that one of the leads was not fully inserted into the ipg header.Physician was able to clean off the lead and reinsert.Mri mode was turned on.This addressed the issue.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not accessible for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer reference number: 1627487-2019-12837.Related manufacturer reference number: 1627487-2019-12838.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ 7 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9446297
MDR Text Key170121587
Report Number3006705815-2019-04841
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020222
UDI-Public05415067020222
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/18/2020
Device Model Number3662
Device Lot NumberA000072034
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/16/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS LEAD X 2
Patient Outcome(s) Other;
Patient Age36 YR
Patient Weight100
-
-