Model Number 24691 |
Device Problem
Material Rupture (1546)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/29/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).
|
|
Event Description
|
It was reported that balloon rupture occurred.A 1.5mm x 20mm x 143cm coyote es balloon was advanced for dilation.However, during first inflation at 4 atmospheres, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported.
|
|
Event Description
|
It was reported that balloon rupture occurred.A 1.5mm x 20mm x 143cm coyote es balloon was advanced for dilation.However, during first inflation at 4 atmospheres, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported.
|
|
Manufacturer Narrative
|
(b)(6).Returned product consisted of a coyote es balloon catheter.There was blood in the inflation lumen.Microscopic examination revealed a torn a length of 3mm on distal end of the balloon.Microscopic examination of the balloon presented no irregularities in the balloon material or the ro marker that could have contributed to the damage.There was no other damage revealed.The damage to the balloon is consistent with interaction with the lesion.There were numerous kinks throughout the catheter.
|
|